产品展示
优质供应
详细内容
Advanced Operating Modes
Normal
Peak tension and compression
Percentage or sharp break detection
Load comparisons and load averaging
Contact closure
Load limits
Pass-fail limits
Statistical Calculations
Mean with maximum and minimum values
Coefficient of variation with mean and standard deviation
Standard deviation with variance and mean
Percent differentiation
Integral Loadcells
Accuracy ±0.1% full scale
Mechanical overload protection to 150% full scale
Simple Operator Interface
High resolution display
Menus, prompts and function keys for easy use
Standard Outputs - RS-232 serial data, Mitutoyo Digimatic, ±2Vdc Analog
NIST Calibration with Data - available IEC/ISO17025 cert with uncertainty
2 year warranty
Model ounce gram pound kilogram Newton
DFS-250G 8 x 0.004 250 x 0.05 0.5 x 0.0001 - 2.5 x 0.0005
DFS-002 32 x 0.004 1000 x 0.1 2 x 0.002 1 x 0.0001 10 x 0.001
DFS-010 160 x 0.02 5000 x 0.5 10 x 0.001 5 x 0.0005 50 x 0.005
DFS-025 400 x 0.05 10000 x 1 25 x 0.002 10 x 0.001 100 x 0.01
DFS-050 800 x 0.1 25000 x 2 50 x 0.005 25 x 0.002 250 x 0.02
DFS-100 1600 x 0.2 50000 x 5 100 x 0.01 50 x 0.005 500 x 0.05
DFS-200 3200 x 0.4 - 200 x 0.02 100 x 0.01 1000 x 0.1
DFS-500 8000 x 1 - 500 x 0.05 250 x 0.02 2500 x 0.2
Specifications
Accuracy: ±0.1% of full scale
Certification: calibration with NIST data, IEC/ISO17025 optional
Data Sampling Rate: 5000 Hz
Peak Capture Rate: 5000 Hz
Display Update Rate: 10 Hz
Tare Capacity: 10% full scale
Overload Protection: 150% full scale
Display Characteristics: high resolution, dot matrix LCD, 8 lines, 40 characters, adjustable contrast, invert and "hide" capability
Automatic Shut Down: configurable time. May be disabled.
Data Storage: 10 results, optional NEXYGEN software for unlimited storage and automated testing and analysis.
Test Stand Control: Compatible with Chatillon TCD series test stands
Outputs: RS-232, Mitutoyo (Digimatic) and ±2Vdc analog
Power: Battery or direct AC operation. Universal Power: 120V / 230V, Rechargeable nickel metal hydride (supplied)
Battery Life: 30 hours, continuous use. Gauge may be operated with direct AC power source
Instrument Weight: 1.5 lbs
Operating Temperature: 40 to 100 F (4 to 38 C)