产品展示
UV-VIS Variable Wavelength detector
点击次数:1203发布时间:2009/11/2 0:00:00
更新日期:2012/3/8 16:46:51
所 在 地:Czech
产品型号:SAPPHIRE 800
优质供应
详细内容
sapphire 800 uv-vis variable wavelength detector
is a detector with continuously variable wavelength in the range of 190 – 800 nm and a noise level ± 0.5 x 10-5 au.the unit may be used for routine analytical chromatography as well as for preparative chromatography, according to the se-lected cell.the unit is easy to operate, reference and sample signals are available for detector diagnostics as well as information on lamp operating hours.output signal is available in both digital and analogue form on the output connector. it is possible to control the unit using the keypad or an rs232 interface connecting capillaries are led from the detector’s right side panel.the detector is equipped with a high standard deuterium lamp in a special socket that enables easy changing. for visible part of spectrum is used as secondary light source a tungsten lamp. wavelength can be programmatically changed during the performance of an analysis. the detector performs automatic wavelength calibration after the lamp has been switched on. a cell is included as a component of the detector.
specification
wavelength range | 190 - 800 nm |
spectral half-width | 6 nm |
accuracy of adjustment | ± 1 nm |
reproducibility | ± 0.5 nm |
light source | deuterium discharge lamp and tungsten lamp |
noise level at empty cell (240 nm, tc 1 s) | ± 0,5 x 10-5 au |
drift at empty cell (240 nm after 1 h) | 1x 10-4 au/hr |
materials in contact with mobile phase | ptfe; fused silica, stainless steel, vespel, peek |
time constant | fast, middle and slow |
programming | 3 change of wavelength during analysis |
adjustable offset | ± 50 mv |
output for integrator | 1 v/au |
interface | rs 232 |
power supply | 100, 115 and 230, 50/ 60 hz) |
power input | 80 va |
dimensions (w x h x d) | 220 x 170 x 450 mm |
weight | 11 kg |
used cells
standard analytical measuring cell hplc 04 supplied with the unit
volume | 10 µl |
optical path | 5 mm |
diameters of capillaries | input 1/ 16“ x 0,2mm ; output 1/ 16“ x 0,5 mm |
mlcc 02 microcell (optional)
volume | 0,5 µl |
optical path | 0,8 mm |
diameters of capillaries | input 1/ 16“ x 0,2mm output 1/ 16“ x 0,2 mm |
plcc 04/05 preparative cell (optional)
volume | (adjustable) 45 μl; 55 μl; 70 μl |
optical path | (adjustable)0.3 mm; 1.4 mm; 2.4 mm |
connecting of plcc 04 | ss capillaries 1/16" x 1,0mm |
connecting of plcc 05 | fep 1/8“ x 1/16“ |