sapphire 800 uv-vis variable wavelength detector
is a detector with continuously variable wavelength in the range of 190 – 800 nm and a noise level ± 0.5 x 10-5 au.the unit may be used for routine analytical chromatography as well as for preparative chromatography, according to the se-lected cell.the unit is easy to operate, reference and sample signals are available for detector diagnostics as well as information on lamp operating hours.output signal is available in both digital and analogue form on the output connector. it is possible to control the unit using the keypad or an rs232 interface connecting capillaries are led from the detector’s right side panel.the detector is equipped with a high standard deuterium lamp in a special socket that enables easy changing. for visible part of spectrum is used as secondary light source a tungsten lamp. wavelength can be programmatically changed during the performance of an analysis. the detector performs automatic wavelength calibration after the lamp has been switched on. a cell is included as a component of the detector.
wavelength range | 190 - 800 nm |
spectral half-width | 6 nm |
accuracy of adjustment | ± 1 nm |
reproducibility | ± 0.5 nm |
light source | deuterium discharge lamp and tungsten lamp |
noise level at empty cell (240 nm, tc 1 s) | ± 0,5 x 10-5 au |
drift at empty cell (240 nm after 1 h) | 1x 10-4 au/hr |
materials in contact with mobile phase | ptfe; fused silica, stainless steel, vespel, peek |
time constant | fast, middle and slow |
programming | 3 change of wavelength during analysis |
adjustable offset | ± 50 mv |
output for integrator | 1 v/au |
interface | rs 232 |
power supply | 100, 115 and 230, 50/ 60 hz) |
power input | 80 va |
dimensions (w x h x d) | 220 x 170 x 450 mm |
weight | 11 kg |
standard analytical measuring cell hplc 04 supplied with the unit
volume | 10 µl |
optical path | 5 mm |
diameters of capillaries | input 1/ 16“ x 0,2mm ; output 1/ 16“ x 0,5 mm |
mlcc 02 microcell (optional)
volume | 0,5 µl |
optical path | 0,8 mm |
diameters of capillaries | input 1/ 16“ x 0,2mm output 1/ 16“ x 0,2 mm |
plcc 04/05 preparative cell (optional)
volume | (adjustable) 45 μl; 55 μl; 70 μl |
optical path | (adjustable)0.3 mm; 1.4 mm; 2.4 mm |
connecting of plcc 04 | ss capillaries 1/16" x 1,0mm |
connecting of plcc 05 | fep 1/8“ x 1/16“ |
① 凡本网注明"来源:易推广"的所有作品,版权均属于易推广,未经本网授权不得转载、摘编或利用其它方式使用。已获本网授权的作品,应在授权范围内
使用,并注明"来源:易推广"。违者本网将追究相关法律责任。② 本网凡注明"来源:xxx(非本网)"的作品,均转载自其它媒体,转载目的在于传递更多信息,并不代表本网赞同其观点和对其真实性负责,且不承担此 类作品侵权行为的直接责任及连带责任。如其他媒体、网站或个人从本网下载使用 ,必须保留本网注明的"稿件来源",并自负版权等法律责任。
③ 如涉及作品内容、版权等问题,请在作品发表之日起两周内与本网联系,否则视为放弃相关权利。
④易推广页面显示产品信息均由企业自主发布,信息内容真实性、准确性与合法性由相关企业负责,易推广对此不承担任何责任,如遇非法或侵权信息欢迎监督,请联系QQ:1273397930或者发邮件至:1273397930@qq.com,如有确实证件证明属实,本站将对其删除处理,谢谢!
⑤ 本信息由注册会员:Shantou Keyi instrument & Equipment co.ltd发布并且负责版权等法律责任。
易推广客服微信