您的位置:首页 > 产品展厅 > 仪器仪表 > 流量仪表 > 其他气体分析仪 > 三氯甲烷/氯仿气体检测管
三氯甲烷/氯仿气体检测管137LA 检测范围0.5-30ppm 三氯甲烷/氯仿气体检测管137LA 检测范围2-12ppm
Measuring Range
0.5to2ppm 2to12ppm 12to30ppm Number of Pump Strokes
4 2 1 Correction Factor
0.25 1 2.5 Sampling Time 2 minutes per pump stroke Detecting Limit 0.2 ppm (n=4) Colour Change White → Pale purple Reaction Principle CHCl3 + I2O5 + H2S2O7 → Cl2
Cl2 + (CH3C10H5NH2)2→ Pale purple products Coefficient of Variation 10% (for 2 to 4 ppm), 5% (for 4 to 12 ppm) Shelf Life 1 Year Corrections for temperature & humidity Temperature correction is necessary Store the tubes in the refrigerator to keep at 10ºC (50ºF) or below.
Substance Concentration Interference Change colour by itself to
Chlorine No No Bromine No No Iodine No No Unsaturated halogenated HCs + Pale purple Saturated halogenated HCs + Pale purple
Calibration gas generation Diffusion tube method
TLV-TWA TLV-STEL Explosive range 10 ppm - -
目盛範囲: 2~12ppm〔基準n=2〕 測定範囲: 0.5〔n=4〕~30ppm〔n=1〕 吸引回数〔n〕: 4,2〔基準〕,1 1回(100ml)の吸引時間: 2分 検知限度: 0.2ppm〔n=4〕 変色: 白色→淡紫色 温?湿度補正: 温度 有効期限: 1年(冷蔵庫保存) 指示精度: CV=5~10%
(CV:変動係数=σ:標準偏差÷平均値×100)
ガス名 共存の場合の濃度と結果 単独の場合
塩素 影響しない 変色しない 臭素 影響しない 変色しない ヨウ素 影響しない 変色しない 不飽和ハロゲン化炭化水素 +誤差 淡紫色に変色 飽和ハロゲン化炭化水素 +誤差 淡紫色に変色
校正用ガス ガス拡散管法
許容濃度 管理濃度 TWA STEL 爆発範囲 3ppm 3ppm 10ppm -- --
① 凡本网注明"来源:易推广"的所有作品,版权均属于易推广,未经本网授权不得转载、摘编或利用其它方式使用。已获本网授权的作品,应在授权范围内
使用,并注明"来源:易推广"。违者本网将追究相关法律责任。② 本网凡注明"来源:xxx(非本网)"的作品,均转载自其它媒体,转载目的在于传递更多信息,并不代表本网赞同其观点和对其真实性负责,且不承担此 类作品侵权行为的直接责任及连带责任。如其他媒体、网站或个人从本网下载使用 ,必须保留本网注明的"稿件来源",并自负版权等法律责任。
③ 如涉及作品内容、版权等问题,请在作品发表之日起两周内与本网联系,否则视为放弃相关权利。